PowerPac™HV Power SupplyInstruction ManualCatalog Numbers164-5056and164-5059

Página 2

21.2 FeaturesThe PowerPac HV has the following features:• Constant voltage, voltage ramping, constant current or constant power operation withautomat

Página 3 - Technical Support

1.3 UnpackingWhen the power supply is received, carefully inspect the container for any damage whichmay have occurred in shipping. Severe damage to t

Página 4

Section 2Control PanelFig. 4. Front Panel.Key DescriptionRun/Pause Starts or pauses a run. When paused, the run parameters for the current or subsequ

Página 5

Section 3Product Setup3.1 Power Supply ConnectionsThis section describes how the PowerPac HV power supply is connected to an electrophoresiscell or c

Página 6

Step Procedure Description3. Turn power on Use the switch located on the right side of the powersupply to power on. Initially, the unit will display

Página 7 - Table of Contents

3.3 Explanation of Safety Features and Setup Options3.3.1 Pfd: Power Failure DetectionPreference Options Description PFd OFF OFF: Turns power failur

Página 8 - List of Figures

3.3.2 RRCD: Rapid Resistance Change DetectionPreference Options Description RRCd ON ON: Detection of a sudden resistance Rapid Resistance Change (D

Página 9 - Introduction

KEY CHIRP ON Activates the audible signal for keystroke.(Default)OFFDATE MM/DD/YY This is the only format available.Press the change soft key then e

Página 10 - 1.2 Features

Section 4OperationThe PowerPac HV has four operation modes:1. Manual: Single-step run, not stored in memory. (see Section 4.1)2. Basic Methods: Nine (

Página 11 - 1.3 Unpacking

4.1 Manual ModeManual mode allows the user to program and start a single-step method with a minimumnumber of keystrokes. This single-step method is n

Página 13 - Product Setup

4.1.1 Create, Edit and Run in Manual ModeStep Procedure Description1. Set up the power Refer to Section 3 for instrument setup.supply2. Select the mo

Página 14 - 3.2 Preferences Setup

Step Procedure Description6. Start the run Press to start a run. A run may be started fromanyscreen of the EDIT loop as long as a valid constant para

Página 15

4.2 Basic Methods ModeBasic Methods Mode is used to create, run and store basic methods. It has the same programming options (edit loop) as the Manua

Página 16

4.2.2 Create, Edit and Run a Basic MethodThis section describes how to create, open, modify and run Basic Methods.Step Procedure Description1. Set up

Página 17 - IRDA ON

Step Procedure Description5. Method steps Press EDIT to enter the edit loop.New method • Press the up and down arrows to select an existing step or se

Página 18 - Operation

Step Procedure Description8. Select the time mode Use the soft keys to select a time mode and thealphanumeric keypad to enter the run time.HRS: Run t

Página 19 - 100 V 500 mA

4.3 IEF Methods ModeThe IEF Method mode is used to create, run and store Isoelectric focusing (IEF) methods.The method list stores up to nine methods

Página 20 - AND VHOURS

4.3.1 Create, Edit, and Run an IEF MethodThis section describes how to create and edit methods.Step Procedure Description1. Set up the power supply

Página 21

Step Procedure Description4. Method name Name a method.• Press the up arrow key to select the methodname field.• Use the CE button to clear the name

Página 22 - 216 V 235 mA

Step Procedure Description8. Enter parameter limits Current or power limits must be set according to application instructions to prevent overheating

Página 23

Technical SupportFor Technical Support contact your local Bio-Rad office. In the US call 1-800-424-6723.Please have the following information availabl

Página 24 - CONSTV LIMITA LIMITW

Step Procedure Description13. The Run Completed or Run Terminated screen isdisplayed at the end of a run. Press RERUN torepeat the run or EXIT to disp

Página 25 - HRS VHOURS UNTIMED

4.4.1 Create, Edit and Run in Temperature MethodStep Procedure Description1. Set up the power supply Refer to Section 3 for instrument setup.2. Conn

Página 26 - 2000 V 0.00 mA

Step Procedure Description7. Select the time mode Use the soft keys to select a time mode and thealphanumeric keypad to enter the run time.HRS: Run

Página 27

4.5 Edit a Paused RunThe PowerPac HV permits editing the run parameters during a pause in all operation methods modes: Manual, Basic, IEF Methods, an

Página 28 - CANCEL OK

4.5.1 Editing a Paused RunStep Procedure Description1. Pause the run Press the PAUSE soft key or to pause a runthat is in progress. The S#/#indicat

Página 29

Section 5Data Transfer Using the Infrared (IR) Port and Data TransferSoftware5.1 OverviewThe PowerPac HV is capable of transmitting methods and data

Página 30 - 216 V 235 mA

5.2 System RequirementsMinimum PC specifications• Windows 2000 or Windows XP Operating System• 400 MHz processor• 256 MB RAM minimum, 512 MB recommen

Página 31

Section 6Maintenance and Troubleshooting6.1 MaintenanceThe PowerPac HV requires little maintenance to ensure reliable operation. To clean thecase, fi

Página 32

6.1.2 Replacement PartsCatalogNumber Description900-8933 Replacement Fuse, 8 A, 250 V, 5 x 20 mm, type T165-9005 Power Cable165-5058 Temperature Pro

Página 33 - 216 V 235 mA

6.3 Error MessagesError Message Possible Cause SolutionERROR STOP: CODE 01 1. Electrophoresis cell 1. Make sure all electrical NOLOAD DETECTED no

Página 35 - Software

Error Message Possible Cause SolutionERROR STOP: CODE 09 1. Loose output 1. Verify all electrical RAPID CHANGE IN R connections leading connectio

Página 36 - 5.2 System Requirements

Error Message Possible Cause SolutionERROR STOP: CODE 18 The set temperature cannot TEMP REG ERROR be reached because:CORRECT AND CONTINUE 1. The t

Página 37 - Section 6

Appendix ATechnical SpecificationsOutput (programmable)Voltage 20–5,000 VCurrent 0.05–500 mAPower (maximum) 1–400 WOutput regulation Constant volta

Página 38 - 6.2 Troubleshooting

Appendix BWarranty and Ordering InformationWarrantyThe PowerPac HV power supply is covered by a standard Bio-Rad Laboratories warranty.Contact your lo

Página 43 - Appendix B

Life ScienceGroup Sig 121210002468 Rev D US/EGBio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 424 6723 Australia 61 2 9914 2800 A

Página 44

SafetyRead this page before proceeding.Bio-Rad PowerPac power supplies are designed and certified to meet EN 61010* safetystandards. Certified product

Página 46

Table of ContentsPageSection 1 Introduction ...11.1 Overview...

Página 47

List of FiguresFigure 1. Front View.......1Figure 2. Rear Vie

Página 48 - Laboratories, Inc

Section 1Introduction1.1 OverviewThe PowerPac HV power supply is designed to provide constant voltage, current, or powerfor a wide range of electroph