Microplate Manager User Guide
92
Control Report
Figure 67. Control Report.
The Control Report displays absorbance or velocity data for each well as a
percentage of the range between the positive and negative controls in the assay.
Controls are defined in templates. For more information about controls, see section
4.3.
If no positive control is defined for the assay, the highest absorbance value or highest
velocity among the wells defined as unknowns or standards is used. If no negative
control is defined, then 0.000 is used as the negative control.
½ Controls are different than limits. Any limits you set will not affect this report.