Bio-rad S3™ Cell Sorter Manual do Utilizador Página 1

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Bio-rad S3™ Cell Sorter. Bio-Rad S3™ Cell Sorter User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - Biosafety System Class I

S3™ Biosafety System Class IInstruction ManualCatalog #145-1078 #145-1021 #145-1022#145-1029 #145-1030

Página 2

viii | S3 Biosafety System

Página 3 - Bio-Rad Technical Support

Instruction Manual | 1Introduction1The S3™ biosafety system is a Class I aerosol containment hood designed to work with the S3 and S3e cell sorters, w

Página 4

IntroductionSystem ComponentsThe S3 biosafety system includes the following components (Figure 1): nEnclosure top assembly with vinyl side panels nHDP

Página 5 - Instrument Warning Labels

Hardware OverviewHardware OverviewThe S3 biosafety system comprises three components (Figures 2 and 3): n HEPA air generator — fan system will filter a

Página 6 - Biohazards

Introduction4 | S3 Biosafety SystemThe front panel of the S3 biosafety system has two access doors (Figure 4). n Sample access door — provides acce

Página 7

Instruction Manual | 5The S3™ biosafety system is a Class I system designed to protect the user and the environment from aerosols generated during the

Página 8

6 | S3 Biosafety SystemOperation IMPORTANT! If the airflow through the HEPA filter falls below required limits for containment, the software will aler

Página 9 - Table of Contents

Instruction Manual | 7Software ControlIf any of the following occur, the S3 or S3e cell sorter will stop sample acquisition or sort and alert the user

Página 10

Operation8 | S3 Biosafety SystemWorking with the S3 and S3e Cell SortersThe S3 biosafety system was custom-designed to provide an easy access interf

Página 11 - Introduction

Working with the S3 and S3e Cell SortersInstruction Manual | 9Fluidics Access DoorThe fluidics access door is positioned for easy access to the bulk fl

Página 13 - Hardware Overview

10 | S3 Biosafety SystemOperationVinyl Side Panel OperationAll four sides of the S3 biosafety system can be opened to access the S3 or S3e cell sort

Página 14 - Specifications

Instruction Manual | 113The S3™ biosafety system has been designed to maintain the safety of the user and the outside environment. Always follow the p

Página 15 - Operation

12 | S3 Biosafety SystemMaintenanceFilter ReplacementReplacement of the HEPA filter requires proper training and should be performed by the certifyin

Página 16 - Software Control

Instruction Manual | 13Whenever service is conducted on the biosafety system, be sure to completely power down both the S3™ or S3e cell sorter throug

Página 17 - Instruction Manual

Service and CertificationCertification ProcedureGeneral InformationThe enclosure is a ventilated cabinet for user and environmental protection. All in

Página 18 - Using the Access Doors

Certification Procedure3. Inject PAO in such a manner to allow for uniform particle distribution across the entire face of the HEPA filter.4. Set the

Página 19 - Fluidics Access Door

Service and Certification16 | S3 Biosafety SystemS3 Biosafety System Class I Certification FormS3 Biosafety System Class I Certification Form Custome

Página 20 - Vinyl Side Panel Operation

Contaminant TestingInstruction Manual | 17Contaminant Testing This protocol is designed to test the aerosol contamination levels that occur under norm

Página 21 - Maintenance

Service and Certification18 | S3 Biosafety SystemTo start sorting the Glo Germ particles:12. Click Start on the instrument control panel to acquire

Página 22 - Filter Replacement

Contaminant TestingInstruction Manual | 19Sample the following locations for 10 min each: nNear sample station nFront, below bulk fluids door nOn top,

Página 23 - Service and Certification

Instruction Manual | iCopyright © 2014 by Bio-Rad Laboratories, Inc. All rights reserved. Toll Free: 1-800-4-BIORAD (1-800-424-6723)Fax: 510-741-5800F

Página 24 - Certification Procedure

20 | S3 Biosafety System Contaminant Testing RecordCustomer Name

Página 25 - References

Instruction Manual | 21Troubleshooting5Table 7 describes possible errors and causes with troubleshooting steps. Table 7. Troubleshooting table. Error

Página 26 - 16

22 | S3 Biosafety System

Página 27 - Contaminant Testing

Ordering Information Catalog # Description 145-1078 S3 Biosafety System Class I145-1021 S3 Cell Sorter 488 1L2D with S3 Biosafety System Class I14

Página 30 - Contaminant Testing Record

Life ScienceGroup Sig 121310033232 Rev B Bio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 424 6723 Australia 61 2 9914 2800 Austri

Página 31 - Troubleshooting

ii | S3 Biosafety SystemBio-Rad Laboratories ResourcesBio-Rad provides many resources for scientists. Table 1 lists available resources and how to l

Página 32 - 22

Instruction Manual | iiiSafety and Regulatory ComplianceFor safe operation of the S3 biosafety system with use of the S3 or S3e cell sorter, we strong

Página 33 - Ordering Information

iv | S3 Biosafety SystemSafe Use Specifications and ComplianceElectrical Safety Information and ClassificationThe S3 biosafety system conforms to inte

Página 34

Instruction Manual | vChemical HazardsThe S3 or S3e cell sorter inside the S3 biosafety system includes a waste container that may potentially contain

Página 35

vi | S3 Biosafety System

Página 36 - Laboratories, Inc

Instruction Manual | viiTable of ContentsChapter 1 Introduction ............ 1System Components ......

Modelos relacionados S3e™ Cell Sorter

Comentários a estes Manuais

Sem comentários