Bio-rad Model 583 Gel Drying Supports Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Acessórios para água Bio-rad Model 583 Gel Drying Supports. Bio-Rad Model 583 Gel Drying Supports User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - HYDROTECH™ VACUUM PUMP

INSTRUCTION MANUALCatalog Numbers165-1781165-1782165-1789165-1790HYDROTECH™ VACUUM PUMPHydroTech Vacuum Pump

Página 2 - TABLE OF CONTENTS

3.0 MAINTENANCE AND TROUBLESHOOTING3.1 MAINTENANCEThe HydroTech pump’s reservoir water should be changed at least once a month or if frequent use of t

Página 3 - Caution/Warning

3.2 TROUBLESHOOTING GUIDERefer to Table 6, Troubleshooting Guide, for HydroTech Vacuum Pump troubleshooting. For further assis-tance, call 1-800-4BIO

Página 4 - 1.0 INTRODUCTION

APPENDIX A. APPROXIMATE GEL DRYING TIMESThe following table provides approximate drying times for various gel concentrations, given the gel dryer’stem

Página 5 - 1.2 PHYSICAL DESCRIPTION

APPENDIX B. SPECIFICATIONSVacuumStarting > 29” Hg (23.4 Torr)After 10 hours > 28” Hg at 22° C (48.8 Torr)At 35° C RT > 26” Hg at start (99.6

Página 6 - Rear View Physical Features

APPENDIX C. WARRANTY AND ORDERING INFORMATIONThe HydroTech Vacuum Pump is warranted for 1 year against defects in materials and workmanship. If anyde

Página 7 - 2.2 SET UP

WARRANTY INFORMATIONModel: ______________________________________________________________________Serial Number:_______________________________________

Página 8 - HydroTech Set Up

Life ScienceGroupWeb site www.bio-rad.com Bio-Rad Laboratories Main Office 2000 Alfred Nobel Drive, Hercules, CA 94547, Ph. (510) 741-1000, Fx. (510)

Página 9 - 2.3 BASIC OPERATION

Table of ContentsTABLE OF CONTENTSSafety ...

Página 10 - 3.1 MAINTENANCE

SafetySAFETYCaution/WarningNo user-serviceable parts are inside. Refer servicing to Bio-Rad service personnel. This instrument is intended for labor

Página 11 - 3.2 TROUBLESHOOTING GUIDE

1.0 INTRODUCTIONBio-Rad’s HydroTech Vacuum Pump is designed for use with the Model 583 Gel Dryer or any other conven-tional slab gel dryer. It uses w

Página 12 - Table 7

1.2 PHYSICAL DESCRIPTIONTable 1.Front View Physical FeaturesFeature Description1. Vacuum Inlet Quick disconnect valve for connecting to the gel dryer’

Página 13 - APPENDIX B. SPECIFICATIONS

Table 2.Rear View Physical FeaturesFeature Description1. Reservoir Storage tank for recirculating water.2. Reservoir Cap Screw-on cap for water reserv

Página 14

2.0 SET UP AND BASIC OPERATIONThe HydroTech Vacuum Pump is shipped fully assembled. To use the pump, simply fill the reservoir withwater and attach t

Página 15 - ORDERING INFORMATION

Table 3 (continued)HydroTech Set UpStep Procedure Description2. Connect the AC power cord. Plug the AC power cord into the HydroTech power inputmodule

Página 16 - Laboratories

2.3 BASIC OPERATIONWarningTurn the vacuum line’s two-way valve to the closed position when the dryer is not in use; the vacuum pump then can continue

Comentários a estes Manuais