ProteinChip® SEND ID ArrayInstruction ManualCatalog #C57-30081For Technical Support, contact your local Bio-Rad office, or in the U.S., call 1-800-
1 © 2007 Bio-Rad Laboratories, Inc.Usen Peptide analysisIntroductionFundamental to most applications of laser desorption/ionization mass spectrom
2 © 2007 Bio-Rad Laboratories, Inc.Storage and Packaging Store ProteinChip SEND ID arrays at room temperature. They should be stored in the foil p
3 © 2007 Bio-Rad Laboratories, Inc.Technical Considerationsn If your sample has salt or urea contamination, proceed directly to Protocol 2. Other
4 © 2007 Bio-Rad Laboratories, Inc.Chemical NoiseBecause ProteinChip alpha-cyano-4-hydroxycinnamic acid (CHCA) matrix is integral to the array sur
5 © 2007 Bio-Rad Laboratories, Inc.Protocol 1 The following method can be used to apply a sample directly to the ProteinChip SEND ID array for pe
6 © 2007 Bio-Rad Laboratories, Inc.Ordering InformationCatalog # DescriptionC57-30081 ProteinChip SEND ID Arrays, A–H format, 12C20-10002 Protein
The SELDI process is covered by U.S. patents 5,719,060, 6,225,047, 6,579,719, 6,818,411, and other issued patents and pending applications in the U.S.