Página 1 - Instruction Manual

Prefilled Poly-Prep®Columns for Ion ExchangeChromatographyInstruction ManualLIT84B 9/1/98 8:36 AM Page FC1

Página 2 - Table of Contents

Section 9Ordering InformationCatalog Ionic MeshNumber Column Packing Form Size731-6211 AG 1-X8 Resin, 50 Cl–100-200731-6212 AG 1-X8 Resin, 50 Cl–200-4

Página 3 - Introduction

Bio-Rad Laboratories, 2000 Alfred Nobel Dr., Hercules, CA 94547LIT84 Rev BLIT84B 9/1/98 8:37 AM Page BC1

Página 4 - Instructions for Use

Table of ContentsSection 1 Introduction ...1Section 2 Column Description...2Section 3 Instructions For Use...

Página 5 - Ion Exchange Capacity

Section 1IntroductionPrefilled Poly-Prep columns are smallpolypropylene columns containing ionexchange resins. These columns combine theanalytical qua

Página 6 - Flow Rate Characteristics

Poly-Prep columns are ideal for routineapplications in which a small amount of chro-matographic medium is used and then dis-carded, or for multiple sa

Página 7 - Chemical Compatibility

2. Remove the column cap to allow the eluantto flow freely.3. Snap off the seal at the bottom of thecolumn.4. Monitor the level of liquid in the reser

Página 8 - Troubleshooting

The ion exchange capacity, usuallyexpressed as meq/ml of resin, is equivalent tothe number of ion exchange sites per unit vol-ume of resin. The follow

Página 9

Section 6Chemical CompatibilityTable 2 gives the chemical compatibilityof prefilled Poly-Prep columns. While the col-umn is often very resistant, the

Página 10 - Ordering Information

Section 7Temperature LimitsIncreases in temperature tend to decreaseresin selectivity, and can be used to decreaseelution times. Increased temperature

Página 11 - LIT84 Rev B

can be removed easily by rinsing with 2-3bed volumes of mobile phase.3. Effluent is red: After prolonged storage,sulfonic acid will sometimes leach fr