Bio-rad BioLogic LP System Manual do Utilizador Página 13

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 28
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 12
Below are the recommended tubing lengths when installing the instrument:
SV-3 to Waste [31 inches (78 cm)]
Pump Outlet to MV-6 [22 inches (57 cm)]
SV-3 to Fraction Collector -or- Conductivity Cell to BioFrac [25 inches
(63 cm)]
MV-6 to Waste [33 inches (84 cm)]
Sample Loop [23 inches (59 cm)]
Mixer to Buffer A [26 inches (67 cm)]
Mixer to Buffer B [26 inches (67 cm)]
MV-6 to Column Inlet [8 inches (21 cm)]
Mixer to Pump Inlet [22 inches (55 cm)]
Conductivity Cell to SV-3 [8 inches (20 cm)]
Column Outlet to UV [9 inches (23 cm)]
UV to Conductivity Cell [8 inches (20 cm)]
Section 12
Connecting Cartridge and Sample Loop
1. Bio-Scale
Mini cartridges CAN be installed in the same holder that
supports the conductivity flow cell (shown in Figure 6).
Fig. 6.
9
4006073D.qxp 8/17/2009 7:45 AM Page 9
Vista de página 12
1 2 ... 8 9 10 11 12 13 14 15 16 17 18 ... 27 28

Comentários a estes Manuais

Sem comentários