5 | Method Editor
130 | NGC Chromatography Systems and ChromLab Software
2. Provide the requisite information for the column. A red asterisk (*) indicates the
field is required.
Note: The Column Name field has a 45-character limit.
The Column Volume, Medium Bed Height, and Hardware Diameter fields are
connected. When you enter or change values for any two fields, the system
automatically calculates the value for the thir
d field.
The Default Linear Flow Rate and Max Linear Flow Rate fields are read-only.
These fields ar
e automatically calculated when you enter values in the Default
Flow Rate and Max Flow Rate fields, respectively.
3. Click Save.
The new column appears in the Column Type dropdown list under the User
Defined category
.