Bio-rad Model EM-1 Econo™ UV Monitor Manual do Utilizador Página 20

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 19
17
Appendix C
Rear Panel Connections
Both the signal output and the AUX sockets are 8-pin circular
mini-DINs. The following information is provided for those wishing to
interface the Model EM-1 Econo UV Monitor with other equipment
such as integrators and non Bio-Rad fraction collectors. System Cable 7
may be used as a breakout cable, with an 8-pin mini-DIN at one end and
loose wires at the other.
Do not attempt to use the socket for anything other than the
Model EM-1 Econo UV Monitor optics unit.
All digital inputs use TTL circuitry and are active LOW.
Pin # Function Notes
1 Connects to Paper stop
AUX Pin 1
2 No Contact
3 No Contact
4 Integrator (+) 0 to 1 V = 0 to 2 AU
5 Chart Recorder (+) 0 to 1 V = Full scale
6 Connects to Pen down
AUX Pin 6
7 Ground
8 No Contact
Pin # Function Notes
1 Connects to
Pin 1
2 Auto-Zero
3 Large Mark Marks chart recorder
signal only
4 Integrator (+) 0 to 1 V = 0 to 2 AU
5 No Contact
6 Connects to
Pin 6
7 Ground
8 Small Mark Marks chart recorder
signal only
pin
configuration
pin
configuration
M7318160G.qxd 05/08/2003 2:06 PM Page 17