Section 9
Troubleshooting
Problem Possible Cause Solution
Pump displays Power surge or 1. Turn the power switch off.
"OUT" brownout 2. Press and hold down the
down arrow key and the CAL
key simultaneously while turning
the power switch on. This will
reset the pump. If the pump was
previously calibrated you will
have to recalibrate it.
Pump displays Power outage If you wish to continue a run in
"FAIL" progress, press program run key.
Otherwise, press the run/stop
key.
Pump will not run No power Verify that powercord is plugged
(no LEDs lit) in and the power switch is turned
on.
Fuses Check that proper fuses are
installed and intact.
Insure that the calibration indica-
torlight is not flashing. For pump
to run, this light must either be
off or steadily lit.
Liquid remains Platen not adjusted Platen too loose. Turn platen
stationary in tubing screw clockwise just until liquid
begins moving within tubing.
Clogged tubing, Check tubing, fittings, and valves
valves, or fittings for obstructions.
Damaged tubing Check pump tubing for damage.
Replace damaged tubing.
Flow rate not Pump tubing Replace tubing.
consistent damaged or worn out
Platen adjustment Adjust platen.
incorrect
Wrong tubing size Re-select tubing size on front
selected panel of pump.