Screen 6. Selecting Units for Programming.
Programming steps can be described in either units of time (in
minutes) or volume (in ml).
6 Program Steps
TIME vol
Pressing the left soft key selects time; the right key selects volume.
Pressing Next advances to screen 8 if a Splitter valve is present or
screen 9 if a Splitter valve is not present.
Screen 7 (not displayed). This is available for future upgrades.
Screen 8. Splitter Valve Period Selection.
The Econo Gradient Pump automatically detects the presence of a
splitter valve. The default period is 2 seconds. This parameter, in
conjunction with the “%-split” parameter, may be varied for stream
splitting, process sampling, or on-line biological activity assays. More
information is included in the splitter valve instruction manual. If you
are uncertain about this value, accept the default of 2 seconds and
adjust it later according to your application needs.
8 Split period
__ sec
Enter a new value for the split period and press Next to advance to
screen 9.
System Operation
17