40 | S3 and S3e Cell Sorters
Getting Started
4.4.5 Drop Delay Maintenance
After drop delay determination, the system will be automatically placed into a droplet monitor
mode to maintain the break-off during use. The changes required for this are completed
automatically. To verify that the drop delay is being maintained, use the Droplet Monitor button
on the tool bar. This window will note if the system is maintaining (Figure 41).
Fig. 40. Drop Delay Alignment Status window.
Fig. 41. Droplet Monitor window with status
and recalculate button highlighted.
Recalculate drop delay
Verify manually
Accept