BIO RAD 400-0174 Instruções de Operação Página 439

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 485
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 438
Error Messages
Check Remedy
DIOM
Message Meaning Possible cause
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
16024 DIOM DIG3 status Error: Status
16023 DIOM DIG2 status Error: Status
for the valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
16025 DIOM DIG4 status Error: Status
Vista de página 438
1 2 ... 434 435 436 437 438 439 440 441 442 443 444 ... 484 485

Comentários a estes Manuais

Sem comentários