Error Messages
Check Remedy
DIOM
Message Meaning Possible cause
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace coil
Solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Run electronics diagnostics
16072
DIOM PWM11
Error: Over-current
16071 DIOM PWM10
current
Error: Over-current
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
16072
DIOM PWM11
current
Error: Over-current
16073 DIOM PWM12
current
Error: Over-current