Bio-rad ChromLab™ Software Manual do Utilizador Página 132

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 306
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 131
3 | Preparing the Instrument
130 | NGC Chromatography Systems and ChromLab Software
8. Remove the syringe from the priming port and discard the liquid into waste.
9. Repeat steps 4–8 for the remaining buffer inlet ports.
10. To prime system pump B, select buffer B1 and perform steps 3–9 on system
pump B.
To prime the sample pump
1. Immerse the inlet tubing for the sample pump in the appropriate container.
2. Insert a 30 ml syringe into the priming port.
3. Turn the priming port valve counter-clockwise one full turn to open the seal.
4. Slowly withdraw the
plunger to withdraw liquid from the port
until
air no
longer enters the syringe (approximately 10 ml)
and all tubing is filled with
liquid.
5. Turn the priming port valve clockwise to close the port.
6. Remove the syringe from the priming port and discard the liquid into waste.
Vista de página 131
1 2 ... 127 128 129 130 131 132 133 134 135 136 137 ... 305 306

Comentários a estes Manuais

Sem comentários