Bio-rad ChromLab™ Software Manual do Utilizador Página 269

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 306
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 268
Plumbing NGC Systems
Instrument Guide | 267
To plumb and prime the system pump wash lines
1. Mount a tube holder to the right or left side of the instrument in either the upper
or lower rack slots.
2. Fill a 50 ml tube with 20% isopropanol (IPA) and insert it in the tube holder.
Tip: You can use 20% ethanol as an alternative.
3. Connect inlet tubing to the pump wash inlet port (left rear port) on
system pump A.
4. Immerse the other end of the pump A wash inlet tubing in the tube of 20% IPA.
5. Connect the pump head wash priming tubing to the outlet (right rear port) on
system pump B.
6. Connect a 1/4-28 female to female luer adaptor to the pump head wash
priming tubing.
7. Connect a 1/4-28 union to a second inlet tubing. You will use the inlet tubing as
the outlet tubing for system pump B.
Tip: Set the inlet tubing with the attached union aside. Keep it close by, as
you will attach it to the pump head wash priming tubing on step 12.
Vista de página 268
1 2 ... 264 265 266 267 268 269 270 271 272 273 274 ... 305 306

Comentários a estes Manuais

Sem comentários