Bio-rad ChromLab™ Software Manual do Utilizador Página 134

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 133
5 | Plumbing and Priming the NGC Chromatography Systems
132 | NGC Chromatography Systems and ChromLab Software
4. In the fluidic scheme, touch or click the System Pump module to open its
dialog box.
5. In the System Pump dialog box, select Priming on the Mode dropdown list.
Note: When in Priming mode, the sample inject valve switches the system
pump flow to waste (W2). After the pr
ocess is completed, the sample inject
valve switches the system pump to manually load the column through the
loop.
6. Select Q4 as the port to open and prime.
Vista de página 133
1 2 ... 129 130 131 132 133 134 135 136 137 138 139 ... 177 178

Comentários a estes Manuais

Sem comentários