Bio-rad ChromLab™ Software Manual do Utilizador Página 53

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 52
Inserting Additional Modules
Installation Guide | 5 1
Inserting Additional Modules
WARNING! Disconnect power to the NGC instrument before inserting or
removing any module. No user-serviceable parts are inside any component
unless noted in this manual. Refer servicing questions to Bio-Rad service
personnel.
WARNING!
To reduce the chance of liquid seeping into the instrument, all
open bays must be filled with at least an empty module. The NGC instrument
will not operate if it detects a slot without a module inserted.
To insert or reposition modules on the NGC instrument
1. Ensure that the NGC instrument is powered off.
2. Loosen the screws on the front of the module cover or covers.
3. Firmly grasp both captive screws and pull the module forward out of its bay.
The following image shows an empty bay.
4. Store the module and the screws in a safe place for future use.
Vista de página 52
1 2 ... 48 49 50 51 52 53 54 55 56 57 58 ... 177 178

Comentários a estes Manuais

Sem comentários