5 | Plumbing and Priming the NGC Chromatography Systems
136 | NGC Chromatography Systems and ChromLab Software
To prime the sample inlet valve (if available)
1. Immerse the tubing for the inlet valve in the appropriate buffer container.
2. On the ChromLab computer or the touch screen, start ChromLab in Manual
mode.
3. In the fluidic scheme, touch or click the Sample Pump module with the sample
inlet valve to open its dialog box.
4. In the Sample Pump dialog box, select the first sample inlet port to prime.