Firmware Update Failure
Installation Guide | 163
Firmware Update Failure
If the firmware upgrade fails, restart the NGC instrument. The software attempts the
firmware update again.
If the firmware update fails again:
1. Shut down the NGC instrument.
2. Turn off the power at the back of the instrument for 30 seconds.
3. Power on the NGC instrument. The firmware update procedure starts again.