Power Requirements
Installation Guide | 1 7
Power Requirements
The power supply to the NGC system must be stable and within specifications at all
times to ensure optimal operation. The power cable connected to the power inlet
port must be rated for 7 amps or more.
Mains input voltage AC 100–240V; 50–60 Hz
NGC maximum power usage <750 watts
Number of power sockets Minimum of three power sockets:
One socket for the NGC instrument
One socket for the computer running
ChromLab software
One socket for the BioFrac™ fraction collector
A socket for any supported peripheral
instrument that you attach (for example, the
C-96 autosampler)
Type of power sockets IEC type connections, grounded sockets
Table 4. Power requirements for the NGC instrument and the computer
running ChromLab
Parameter Requirement
Tip: The touch screen does not require a
separate power supply.